Vial Leak Test Optimization
Date
Authors
Advisor
Publisher
Polytechnic University of Puerto Rico
Item Type
Article
- Total Views Total Views2
- Total Downloads Total Downloads7
Abstract
This paper describes the testing performed to determine the optimum operational parameters that will be used to operate the Mar-Tre Leak Testerfor the 12Z vial configurations. Based on the results obtained, the system demonstrated to be capable of detecting leaks on the 12Z vial with 99% reliability and 95% confidence. After optimizing the leak testing process, no escapes or false rejections have been detected. Also, the results obtained confirm thatthe system has repeatability and reproducibility capabilities.Key Terms⎯ANOVA, Leak, Optimization,Vials,
Description
Design Project Article for the Graduate Programs at Polytechnic University of Puerto Rico
Keywords
Citation
Chávez Salazar, R. (2021). Vial Leak Test Optimization [Unpublished manuscript]. Graduate School, Polytechnic University of Puerto Rico.