Missing Magnet Defect Rate Improvement Project at the Cement Unit at Stryker Puerto Rico

Date

Publisher

Polytechnic University of Puerto Rico

Item Type

Article
  • Total Views Total Views1
  • Total Downloads Total Downloads3

Abstract

At Stryker Puerto Rico, there was a growing of customer complaints for the new Prep+ Cartridge. The complaints were generated due to units found without a component. This is a magnet that is placed on the back of the product that allows a connection that makes the product work. Without the magnet, the product is useless. Based on the recognized problem in the organization, an analysis with the application of DMAIC was done to identify the root cause. It was found that the cause of the problem was during the assembly process. Nevertheless, some opportunities were also found in the design of the product. Propositions of Quality Control and Fixtures/Equipment improvements for production process were considered to decrease the defect rate for the missing magnet. Key Terms ⎯ DMAIC, Point of Cause, Fixture, Improvement.

Description

Design Project Article for the Graduate Programs at Polytechnic University of Puerto Rico

Keywords

Citation

Ramos Padilla, H. L. (2023). Missing Magnet Defect Rate Improvement Project at the Cement Unit at Stryker Puerto Rico [Unpublished manuscript]. Graduate School, Polytechnic University of Puerto Rico.