Time Optimization of Gas Chromatography-Mass Spectrometry Method for Controlled Substances Analysis

Date

Publisher

Polytechnic University of Puerto Rico

Item Type

Article
  • Total Views Total Views1
  • Total Downloads Total Downloads8

Abstract

An optimization of the gas chromatography- mass spectrometry (GC-MS) method was conducted by establishing a temperature ramp of 150°C, held by 3 min and raising at 12°C/min until reaching 300°C, which from 25.00 min was reduced to 15.50 min the analysis time and identified qualitatively the Caffeine, Cocaine, Delta 9 THC, Oxycodone and Heroin. A validation of the results was performed to determine the precision, instrument detection limit (IDL) and selectivity from six replicates analysis. For the controlled substances identification a 200 ppm QC-Standard Drug mix solution was prepared. A low variability was observed in the retention time of each compound. Instead, the peaks area demonstrates a high variability from the acceptance criteria of RSD, especially the Oxycodone affecting the IDL results. All the compounds showed a good resolution with acceptance criteria of R<2, except Oxycodone. Further measurements will be necessary to acquire more tight values to the acceptances criteria.

Description

Design Project Article for the Graduate Programs at Polytechnic University of Puerto Rico

Keywords

Citation

Espada Colón, C. A. (2016). Time optimization of gas chromatography-mass spectrometry method for controlled substances analysis. Politechne, 18(1), 58-62.