Mitigating Unintended Fiducial Markings in Non-Toric Intraocular Lenses: An Engineering Management Approach

dc.contributor.advisorCruzado Vélez, Héctor J.
dc.contributor.authorCrespo Sánchez, Oziel
dc.date.accessioned2025-08-13T19:15:51Z
dc.date.issued2025
dc.descriptionDesign Project Article for the Graduate Programs at Polytechnic University of Puerto Rico
dc.description.abstractThe Johnson & Johnson Vision plant in Añasco manufactures Non-Toric intraocular lenses (IOLs), which have recently experienced a recurring issue of unintended fiducial markings. These defects, if undetected, compromise product integrity, patient safety, and production efficiency. Root cause analysis determined human errors specifically incorrect machine programming selection as the primary contributor. To address this, an upgraded version of CNC Program Assembler software was developed to enforce correct program usage through barcode scanning, preventing operational oversights. Complementary procedural actions included operator training, standard operating procedure updates, and engineering test validations. A cost benefit analysis revealed substantial material and labor losses due to incorrect production, justifying the proposed corrective measures. The revised system enhances traceability, reduces waste, and ensures consistent quality. Ongoing monitoring and feedback collection are planned to ensure sustainable improvements and compliance. This initiative reflects a strategic integration of risk management, process optimization, and quality assurance to support continuous improvement.
dc.identifier.citationCrespo Sánchez, O. (2025). Mitigating Unintended Fiducial Markings in Non-Toric Intraocular Lenses: An Engineering Management Approach [Unpublished manuscript]. Graduate School, Polytechnic University of Puerto Rico.
dc.identifier.urihttps://hdl.handle.net/20.500.12475/3135
dc.language.isoen
dc.publisherPolytechnic University of Puerto Rico
dc.relation.haspartSan Juan
dc.relation.ispartofEngineering Management Program
dc.relation.ispartofseriesSpring-2025
dc.rights.holderPolytechnic University of Puerto Rico, Graduate School
dc.rights.licenseAll rights reserved
dc.subject.lcshPolytechnic University of Puerto Rico--Graduate students--Research
dc.subject.lcshPolytechnic University of Puerto Rico--Graduate students--Posters
dc.subject.lcshPolytechnic University of Puerto Rico--Subject headings--Unassigned
dc.titleMitigating Unintended Fiducial Markings in Non-Toric Intraocular Lenses: An Engineering Management Approach
dc.typePoster

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
PUPR_CEAH_SJU_SP25_MEM_Oziel Crespo Sanchez_Poster.pdf
Size:
451.62 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
License Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections