Battery Assembly Re-design in a Medical Device Company
Date
Authors
Advisor
Publisher
Polytechnic University of Puerto Rico
Item Type
Article
- Total Views Total Views1
- Total Downloads Total Downloads2
Abstract
Abstract - Today, customer complaints is a strategic business metric for the organizations to measure the customer satisfaction and product’s performance. During YR2011, an adverse trend on complaints per million (CPMs) was observed in a medical device due to loss of power issues. This represents a decrease in customer satisfaction and product’s performance. The intention of this Design Project was to reduce the CPMs by 50%. DMAIC (Define, Measure, Analyze, Improve and Control) methodology was used as a problem solving tool to identify the root causes and improve the product performance. Brainstorming, Failure Analysis, Cause & Effect diagram and Pareto chart were some of the tools used to understand the relationship between the potential causes and determine which contributes significantly to the top offender. Finally, a solution was implemented to address the root cause of the top offender achieving a reduction of CPMs by 81%.
Key Terms - Complaint per Million (CPMs), DMAIC, Error 0010, Medical Device.
Description
Design Project Article for the Graduate Programs at Polytechnic University of Puerto Rico
Keywords
Citation
Ortiz García, A. Y. (2011). Battery assembly re-design in a medical device company [Unpublished manuscript]. Graduate School, Polytechnic University of Puerto Rico.